21. Touch
Stop
The AutomaticQC cartridge simulator is ejected.
22. Manually position the valve in the AutomaticQC cartridge simulator
back to the Start position.
23. If alignment was not achieved, replace the AQC frame. Refer to
Replacing Components.
Wait while the system returns the valve to the correct position.
24. Touch
Continue

Lamp Calibration Test

Use this procedure to test the CO-ox halogen lamp to check proper lamp
operation and associated power driver and control circuitry. This test is
also used to perform a lamp calibration. During the test, the lamp is turned
on using the most recent lamp calibration settings, and a full lamp
calibration and diagnostic test is performed.
Perform this test whenever the lamp is changed or whenever the
NOTE:
lamp is removed and reinstalled.
1. Touch the
2. Touch
Sample Totals
CAUTION:
are at the Diagnostics screen or while you are performing diagnostic tests.
When you exit from the Diagnostics screen, the system may perform an
extended calibration before you can use the system.
3. Touch
Diagnostics
4. Enter the service password, or use the password of the day.
The Diagnostics screen appears.
5. Touch the
All buttons are disabled and the system performs the Lamp Calibration
test.
6. The system turns off the lamp.
252
when the test is complete.
button three times to return to the Analysis screen.
button
Recall
.
.
The system cannot perform routine fluidic activities while you
.
button.
Lamp Cal
Rev. C