Installing The Rear Cover, Cartridge Interface Assembly, And Power Module Assembly - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Installing the Rear Cover, Cartridge Interface Assembly,
and Power Module Assembly
BIOHAZARD:
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Refer to the figures in Removing the Rear Cover and Power Module
Assembly on page 375 while performing this procedure.
Tools and Supplies
Phillips screwdriver, #1
offset Allen wrench, 3/32
1. Complete the required task(s):
If you are reinstalling
the . . .
rear cover and power
module assembly
cartridge interface
assembly
378
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Observe precautions for handling electrostatic-sensitive
Then . . .
Go to step 2.
a. Reconnect the door latch assembly harness (P60).
Ensure that the harness is positioned to the right-hand
side.
b. Reconnect the wash/waste switch harness (P61).
c. Replace the AQC cable (P40) in the central cable
clamp and reconnect.
d. Reconnect the CO-ox cable (P50) and replace it in the
central cable clamp and the left-hand clamp.
e. Reconnect the female outboard cable harness (J30).
f. Reconnect the male outboard cable harness (P20).
g. Position the cartridge interface assembly in the frame
and reconnect the harness OMZ (5 pin) (P2) to the
sample chamber interface assembly.
NOTE:
heater are not bunched or crimped as you attach the
sample chamber interface assembly.
Check that the wires for the connector block and
(Continued)
Rev. C