For Rapidpoint 405 systems an error
has occurred that may be caused by
failing calibrations, and tHb is unlikely
to become available with further
calibrations. However, the problem
may have a different cause which may
be corrected with further intervention.
Analyze has a line through it.*
Probable Cause
If all parameters are available, the
Rapidlink data management system
or the LIS connected to the Rapidpoint
400 series system has turned analysis
functions off for the Rapidpoint 400
series system.
All parameters are not available
because they have failed scheduled
QC analysis or scheduled QC
analysis was not performed.
*
This condition appears for a few seconds when returning to the analysis screen.
The Continue button has a line through it.
Probable Cause
If this button appears on the Status
screen, the system has an error that
must be corrected before you can
continue.
Rev. C
View the Events Log to determine the event causing
the parameter to fail, and follow the appropriate
solutions for that event.
Corrective Action
Perform other tasks, such as recalling results and
accessing the Setup options, if necessary. Contact
your system supervisor for assistance if you need to
analyze samples.
The system is still capable of analyzing QC
samples. Perform the troubleshooting procedures
described in this section for failed QC analysis and
for QC analysis not performed when scheduled.
Corrective Action
Identify and correct the error condition displayed on
the Status screen.
Troubleshooting 159