Replacing the Polychromator Module
BIOHAZARD:
Tools and Supplies
•
Phillips screwdriver, #1
•
small flat-blade screwdriver
•
hex driver
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
3. Remove the cartridge interface assembly. Refer to Replacing the
Cartridge Interface Assembly on page 309.
4. Remove the polychromator module:
a. Disconnect the polychromator fiber bundle from sample chamber
interface assembly.
CAUTION:
the optical fibers resulting in erroneous optics readings.
362
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Do not bend, twist, or crimp the fiber bundle as this can damage
Rev. C