Setting Up Rapidqc Hct In The Rapidlink System; Finishing The Rapidpoint Setup; Swapping Out A Rapidpoint - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

14. After all QC information is entered, select

Setting Up RapidQC Hct in the Rapidlink System

1. Select
Add Control
2. Choose the Rapidpoint 400 name for the new Rapidpoint 400 system
that you just set up.
3. Select
Control Name
4. Verify that all active Parameters are selected.
5. Select
Target Value and Absolute Limits
type.
6. Type in Level names (A and B).
7. Select
Save
8. Select the Rapidpoint 400 system name.
9. Select
Setup Active Levels
10. Select a Level, enter the lot number, the expiration date and start date,
and package insert ranges for each level of Hct QC.
11. Repeat step 10 for both levels of RapidQC Hct.
12. After all QC information is entered, select

Finishing the Rapidpoint Setup

1. At the Rapidpoint 400 series system, enter the Rapidpoint 400 IP
address:
a. Touch
b. Touch
c. Touch
d. Touch
e. Touch either
f.
Enter the Rapidpoint 400 IP address that you copied in step 2e in
Removing the Current Rapidpoint 400 Series System.
2. Touch the
Rev. C
.
(RapidQC Hct).
.
.
Printer and Devices
.
Communications
.
CompleNet
.
Configure
Use DHCP
button twice.
Continue
Swapping Out a Rapidpoint 400 Series System 417
Save
in the Evaluation Technique
Save
.
or
Enter IP Address
.
.
.