Temp Out of Range Descriptions
Temp Out of Range is detected at the end of the sample measurements.
In Retro Calibration Mode
At the end of measurement, the measurement module temperature was
found to be outside the range 37 ± 0.4°C for all sample types.
Not in Retro Calibration Mode
At the end of measurement, the measurement module temperature was
found to be outside the range 37 ± 0.2°C for all patient sample types. For
QC sample types, the range will be 37 ± 0.4°C. This exception is likely
caused by fluid introduction before the instrument had reached the normal
operating temperature.
All Modes
This exception causes the sample results to be aborted.
Refer to the related exceptions: Temp Not Ready and Temp Warning.
224
Rev. C