4. Install a new roll of printer paper:
Refer to Replacing the Printer Paper in Section 3 in the Rapidpoint 400
Series Operator's Manual, if required.
a. Grasp the latch on top of the touch screen and move the screen
forward to expose the printer compartment.
b. Install a new roll of paper. Insert the spindle through the roll of paper
and place the paper in the printer compartment. Ensure that the
paper is tightly wound and the ends of the spindle fit into the
grooves on the sides of the compartment.
When you advance the paper, watch the paper move through the
NOTE:
printer to ensure that it exits the printer correctly.
c. Insert the paper from the bottom of the roll through the back of the
printer.
d. Turn the paper-advance knob clockwise to move 2 to 3 inches of
the paper through the top of the printer.
When you close the printer compartment, ensure that the edge of
NOTE:
the printer paper extends beyond the top of the printer.
e. Close the printer compartment.
f.
Adjust the position of the screen for viewing.
Do not connect the network cable at this time.
NOTE:
5. Connect the power cord to the receptacle on the rear panel of the
system and then to a 100–240VAC, 50–60Hz, 150VA grounded
electrical outlet or to an uninterruptible power supply.
The system automatically detects the voltage when you turn on the
system. Manual setting of the voltage is not necessary.
6. Turn on the power switch, which is located on the rear panel of the
system.
After a few minutes, the system displays the Startup screen. It then
displays the Status screen and a message indicating that the
measurement cartridge needs replacing.
7. Touch
Cancel
392
.