Message
COox Sample
Temp Out of Range
180
Probable Cause and Corrective Action
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
1. Perform the temperature test with the measurement cartridge
installed.
2. Check all cables and connections to and from the sample
chamber interface assembly.
3. Perform the temperature test again.
4. Replace the cartridge. Refer to Replacing the Measurement
and Wash/Waste Cartridges in Section 3 in the Rapidpoint 400
Series Operator's Manual if you want to replace the cartridges.
5. Replace the sample chamber interface assembly.
6. Perform the temperature test again.
7. Replace the connector block.
The system detects that the temperature of the sample is beyond
the acceptable measurement range at the end of sample
analysis. This message usually appears if you analyze a sample
before the CO-ox sample chamber reaches its normal operating
temperature. The system does not report the sample results.
Analyze the sample again. If the message appears again, record
the message and call for technical assistance.
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
1. Perform the temperature test with the measurement cartridge
installed.
2. Check all cables and connections to and from the sample
chamber interface assembly.
3. Perform the temperature test again.
4. Replace the cartridge. Refer to Replacing the Measurement
and Wash/Waste Cartridges in Section 3 in the Rapidpoint 400
Series Operator's Manual if you want to replace the cartridges.
5. Replace the sample chamber interface assembly.
6. Perform the temperature test again.
7. Replace the connector block.
Rev. C