Message
No Paper in Printer
218
Probable Cause and Corrective Action
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
If cartridge replacement does not solve the problem:
Before removing the measurement cartridge,
CAUTION:
ensure that any tests not requiring removal of the
measurement cartridge are performed. Once a measurement
cartridge is removed, if reinstalled it is likely to fail within a few
days. To verify repairs, you may reinstall the measurement
cartridge while remaining in Diagnostics. A new cartridge
must be installed before leaving the customer site.
1. Use the measurement cartridge simulator to perform the
measurement test.
2. Check cables and connections to cartridge interface
assembly.
3. Replace the connector block.
4. Replace the cartridge interface module
The printer is out of paper.
1. Install a new roll of paper as described in Replacing the
Printer Paper in Section 3 in the Rapidpoint 400 Series
Operator's Manual.
2. If necessary, access the Recall menu to locate and print
sample results or calibration data that were not printed while
the printer was out of paper.
Service Solutions
Perform service solutions only after all other corrective actions
have taken place.
1. Perform the printer test.
2. Replace the display printer assembly.
3. Replace the UIP board.
(Continued)
Rev. C