Bayer HealthCare RapidPoint 400 Series Service Manual page 379

Table of Contents

Advertisement

If you are reinstalling
the . . .
2. Reconnect all cables.
a. Attach the fiber bundle to the neon board assembly and the
illumination housing assembly.
b. Replace the bundles in the cable clamp.
3. Reinstall the power module assembly with the 14 screws.
If a CO-ox system, ensure that the fiber bundles are not twisted,
CAUTION:
crimped, or pinched between the frame mounting area as this can result in
erroneous optics readings.
4. Ensure that all cables are not crimped or pinched.
5. Reinstall the rear cover with four Phillips screws: one screw on each
side and two screws on the rear of the cover.
6. Reconnect all external cables.
Rev. C
Then . . .
h. Reconnect the sample chamber interface assembly.
Position the wires for the connect block and heater in
the slot on the side of the sample chamber interface
assembly.
i. Pass the pre-amp cable through the access hole and
reconnect.
j. Reinstall the pre-amp cable cover.
Pull the fiber bundle cables through the thermal
NOTE:
baffle as you install the cartridge interface assembly.
Do not bend, twist, or crimp the fiber bundle as
CAUTION:
this can damage the optical fibers resulting in erroneous
optics readings
k. Reinstall the cartridge interface assembly into the
interface wall.
l. Ensure that the door grounding wire is positioned in the
slot provided.
m. Ensure that the cables are not pinched or crimped and
the clips on the bottom of the frame are engaged to the
interface wall.
n. Replace the two cartridge interface screws.
o. Go to step 2.
Replacing Components 379