3. Reconnect the network cable.
4. If the connection is unsuccessful, shut down and restart the system:
Touch the
NOTE:
Communications screen.
a. Touch the
b. Touch
c. When prompted, turn the power switch off.
d. Wait 10 seconds, and turn the power switch on.
e. Review the setup data entered from step 1 in Finishing the
Rapidpoint Setup and continue with the procedure.
5. If required, enter the FTP Server name or IP address:
a. Touch the
b. Touch
c. Touch
d. Enter the service password.
e. Enter the required information.
6. Repackage the original Rapidpoint 400 series system in the new
Rapidpoint 400 series system's packaging.
7. Attach the return shipment paperwork, included with the swapout
system and ship accordingly.
8. E-mail the files from the Trace Log and Sensor Data diskettes to your
authorized repair depot, or place the packaged diskettes with the
system to be returned.
418
and then touch the
None
button to access the Status screen:
Continue
.
Shutdown
button.
Recall
.
Sample Totals
.
Service Data
button to leave the
Continue
Rev. C

Advertisement

Table of Contents
loading

Related Products for Bayer HealthCare RapidPoint 400 Series

Table of Contents