Qualifier 10 occurs during PUI and some associated faults during normal
operations.
Qualifier 11 occurs any time during data acquisition when the system fails
any tests that check the DAS functionality of the Main board.
Qualifier 13 occurs during washes and pump MITs.
Qualifier 14 occurs if the door opens without the door lock being
deactivated by the system, and if the door fails to open after touching the
Continue button on the message indicating the system is unable to open
the door.