Bayer HealthCare RapidPoint 400 Series Service Manual page 236

Table of Contents

Advertisement

2. Touch
Sample Totals
CAUTION:
are at the Diagnostics screen or while you are performing diagnostic tests.
When you exit from the Diagnostics screen, the system may perform an
extended calibration before you can use the system.
3. Touch
Diagnostics
4. Enter the service password, or use the password of the day.
The Diagnostics screen appears.
5. Touch
M Valve
The second Diagnostic screen appears.
6. Touch the
The system moves the valve.
7. Look for obstructions and listen for any unusual noises that may
indicate a problem with the valve.
If the photosensor detects an obstruction the a D33: 1 error is
NOTE:
posted.
8. Touch
Change
listen for problems.
The valve test moves the valve to three positions numbered 1 through
3. The number of the current position appears on the screen when you
touch
Change
screen.
9. Touch
Stop
Wait while the system returns the valve to the correct position.
10. Touch the
If the diagnostic test indicates a problem(s) with valve operation,
NOTE:
proceed with Using the Measurement Cartridge Simulator on page 240.
236
.
The system cannot perform routine fluidic activities while you
.
.
, and
Start, Change
to move the valve to the next position and then look and
. The test is complete when position 3 appears on the
to stop the test.
button three times to return to the Analysis screen.
Continue
buttons sequentially to run the test
Stop
Rev. C