If the qualifier is . . . Then . . .
2
3
4
7
9
11
12
208
detection occurs during each zero and normal wash when a
repeated optical clarity test BIT failure occurs.
detection occurs during each zero, slope, full calibration, or
CO-ox sample measurement after an analytical detector
signal out of range BIT when the dark measurement exceeds
the light measurement or the light measurement is saturated.
detection occurs during PUI, and for each zero, full
calibration, and CO-ox sample measurement when
excessive noise occurs for the tHb analytical detector during
a dark noise BIT.
detection occurs during PUI, and for each zero, full
calibration, and CO-ox sample measurement when dark
current out of range occurs for the tHb analytical detector
during a dark current BIT.
detection occurs during PUI when an EEPROM alignment
error occurs during the alignment PROM BIT.
detection occurs during each calibration when a neon lamp
failure occurs.
detection occurs during each calibration when the
wavelength shift calculation fails the wavelength shift BIT.