Replacing the Cartridge Interface Assembly

BIOHAZARD:
Observe precautions for handling electrostatic-sensitive
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Tools and Supplies
Phillips screwdriver, #1
small flat-blade screwdriver
Add the label and plug to the front of the cartridge interface
NOTE:
assembly before you begin. See the instructions in the kit.
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover. Refer to Removing the Rear Cover and Power
Module Assembly on page 375.
3. Open the door:
a. Slide the front of the system to the edge of the work surface.
b. Using a small flat-blade screwdriver, access the door lock solenoid
through the middle of the three access holes in the bottom front
right corner of the system and push the solenoid toward the back
of the system until the front door opens.
4. Remove the cartridge interface assembly:
Rev. C
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Replacing Components 309