7. The software loads.
This takes a few minutes, then the system performs a short, simulated
initialization and the clock begins to count down from 20:00 for about
1 minute.
The touch screen displays the Selections screen along with the
message FOR DEMONSTRATION ONLY.
You are now in the FOR DEMONSTRATION ONLY mode.
While in the FOR DEMONSTRATION ONLY mode, do not install
CAUTION:
a real measurement, wash waste, or AutomaticQC cartridge. Real
cartridges become unusable once installed in the FOR DEMONSTRATION
ONLY mode.
8. Use the FOR DEMONSTRATION ONLY mode to practice with setup
options and cartridge replacement.
9. When the demonstration is complete, touch
10. Touch
Replace
The door opens.
11. Remove the demonstration cartridges, including the AutomaticQC
cartridge.
12. Touch the
switch.
422
on the
Cartridge Status
button and when prompted, turn off the power
Shutdown
.
Continue
screen.