No Results on Screens and Reports

If individual results do not appear on screens and reports, ensure that
parameters are not turned off because they failed Required QC or
AutomaticQC or missed Required QC analysis or because they are out of
calibration as described in Button and Parameter Status on page 155.
Refer to this Section to identify other causes of missing results on screens
and reports.
Probable Cause
The parameter was turned off at the
Analysis screen.
The parameter was turned off in
Setup.
The Rapidlink data management
system or the LIS connected to the
Rapidpoint 400 series system turned
the parameter off.
164
Corrective Action
Analyze the sample again, ensuring that the
parameter is turned on at the Analysis screen,
as described in Analyzing Patient Samples in
Section 1 in the Rapidpoint 400 Series
Operator's Manual.
Ensure that the parameter is turned on in
Setup as described in Selecting Parameters
and Units of Measure in Section 6 in the
Rapidpoint 400 Series Operator's Manual.
Contact your system supervisor for
assistance if you need to analyze samples
and obtain results for this parameter.
(Continued)
Rev. C