Circuit Descriptions

Amperometric Isolation Circuits

There are four isolation circuits on the Multiplexer board for the four
amperometric channels (oxygen, glucose, lactate, and creatinine).
Rev. C
Description
Isolated +800 mV reference voltage for the oxygen
electrode.
Isolated -400 mV reference voltage for the glucose
electrode.
For future use.
For future use.
Positive power supply for the potentiometric amplifiers.
Negative power supply for the potentiometric amplifiers.
Power supply for the EEPROM in the sensor module.
Power supply return signal for the EEPROM.
Multiplexed sensor signal, offset adjusted and properly
amplified output of the Multiplexer board.
Signal low sense return for PREMUX_SIG.
Signal connected to the reference side of the bridge.
High terminal of the block thermistor.
Signal connected to the reference side of the bridge.
High terminal of the preheater thermistor.
Signal is not used.
Signal is not used.
Output of the first hematocrit circuit. This signal originates on
the Contact board and passes through the Multiplexer board
on its way to the DAS circuitry.
Output of the second hematocrit circuit. This signal
originates on the Contact board and passes through the
Multiplexer board on its way to the DAS circuitry.