Bayer HealthCare RapidPoint 400 Series Service Manual page 249

Table of Contents

Advertisement

14. Continue to touch
Refer to the following figure which shows the fluid positions on the
AQC valve.
AQC Valve Fluid Positions
The AQC valve positions are tested twice during the alignment test.
NOTE:
Once with the measurement cartridge valve in the wash position and again
with the valve in the ventAQC position. For positions 1 - 7 as indicated on
the display, the measurement cartridge valve is in the wash position. After
position 7 the valve changes to the ventAQC position, and the AQC valve
positions are tested again (positions 8 - 14).
Rev. C
until all fluid positions are observed.
Change
Troubleshooting 249