4. Disconnect the following cables from the UIP board:
UIP display cable
UIP utility cable
Diskette drive
UIP LPT1/comm cable
UIP comm1/inboard cable
5. Remove the three screws securing the UIP board to the standoffs.
6. Remove the UIP board from the Main Interconnect board.
7. Remove the two screws securing the Main Interconnect board to
spacers on the Main board.