7. The system prints the test results as PASS/FAIL:
If . . .
the sample chamber is left in
the open position
the cell remains in the closed
position

Shutting Down the System

Use this procedure to remove power from the system.
To prevent electrical shock or damage to the system, remove power from
the system only as described in this procedure.
1. If prompted, enter your password, the service password, or the
password of the day.
2. Touch the Status button.
3. Touch Shutdown.
CAUTION:
without power. To maintain cartridge stability, do not remove power from
the system for more than 60 minutes if a cartridge is installed.
4. When prompted, turn the power switch off.
The power switch is located on the rear panel of the system.
5. To restore power to the system, turn the power switch on.
After the Rapidpoint 400 series system title screen appears, the Wait
screen displays the time remaining until you can use the system. The
Analysis screen appears when the system is ready to analyze
samples.

Recovering from a Power Loss

Cartridges installed in the system remain stable for 60 minutes without
power. If power is removed from the system for less than 60 minutes, you
can resume operating the system without replacing cartridges.
If power is removed from the system for more than 60 minutes, prepare the
system for use as follows:
256
Cartridges installed in the system remain stable for 60 minutes
Then the system response is . . .
Pass.
Fail. A D73 is posted and CO-ox sample analysis
remains disabled.
Rev. C