c. Replace the four screws.
d. Route the motor wires over the top of the motor and using the small
flat-blade screwdriver, connect it to the outboard cable harness.
Ensure that the wires are not located in any pinch points of the
motor.
Ensure that the pins are seated securely in the outboard cable
harness.
When reinstalling the cartridge interface assembly, do not bump the
NOTE:
spring plate and locking tabs on the pump housing or the spring plate can
pop off and interfere with system operation.
7. Secure the wires with cable ties.
8. Reassemble in the reverse order. Refer to Replacing the Cartridge
Interface Assembly on page 309 and Installing the Rear Cover,
Cartridge Interface Assembly, and Power Module Assembly on
page 378.