Swapping Out a Rapidpoint 400 Series System

This appendix describes the procedure you need to follow when replacing
a Rapidpoint 400 series system, if swapping out is the preferred method of
service.
To perform this procedure, you will need three (3) blank
NOTE:
preformatted, 1.44 MB, 3.5-inch diskettes.
BIOHAZARD:
recommended precautions when working with biohazardous materials.

Removing the Current Rapidpoint 400 Series System

Obtain from the site or create a Setup disk for the institution's current
Rapidpoint 400 series system.
1. Create a Setup disk:
a. If prompted, enter password
b. Touch the
c. Touch
d. Enter the default password,
button.
e. Touch
f.
Touch the down arrow button and then touch
g. When prompted, insert a blank preformatted 1.44 MB, 3.5-inch
diskette into the diskette drive, and then touch the
h. The Wait screen appears while the system copies the data to the
diskette.
i.
When prompted, remove the diskette from the drive and then touch
the
If the current Rapidpoint 400 series system is inaccessible, capture
NOTE:
and record the Rapidpoint 400 series system IP address from the Rapidlink
system. See step 5.
Rev. C
Refer to Protecting Yourself form Biohazards, for
button.
Status
.
Setup
Secured Options
button.
Continue
Swapping Out a Rapidpoint 400 Series System 411
.
12345
, and then touch the
12345
.
Continue
.
Save Setup
button.
Continue