b. Lift the latch.
c. Remove the measurement cartridge.
7. Install the cartridge simulator.
8. Rotate the range switch on the front of the cartridge simulator, to select
the range (low, nominal, or high) for the test. Refer to the following
table:
Measurement
Channel
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Rev. C
Low
Name
Range
GND
-0.25 to +0.25
PO2
-25 to +25
PH
-1382 to -968
GLU_A
-50 to +50
GLU_B
-50 to +50
LAC_A
N/A
LAC_B
N/A
K
-1229 to -809
Na
-787 to -349
Cl
-79 to +388
Ca
-645 to -200
HCO3
-173 to +291
REF
-200 to +200
DAC
N/A
PH2
-1382 to -968
CRE_A
N/A
CRE_B
N/A
FD1/HCT1 -350 to +650
FD2/HCT2 -350 to +650
N/A
N/A
N/A
N/A
N/A
N/A
Nominal
High
Range
Range
-0.25 to +0.25
-0.25 to +0.25
+790 to +970
+2900 to +3600
-119 to +347
+1145 to +1662
-340 to -460
-3000 to -3700
-340 to -460
-3000 to -3700
N/A
N/A
N/A
N/A
-246 to +214
+732 to +1232
+39 to +511
+870 to +1376
+422 to +910
+924 to +1432
+132 to +608
+904 to +1411
+594 to +1089 +1051 to +1882
-200 to +200
-200 to +200
N/A
N/A
-119 to +347
+1145 to +1662
N/A
N/A
N/A
N/A
+750 to +1818 +1850 to +3148
+750 to +1818 +1850 to +3148
N/A
N/A
N/A
N/A
N/A
N/A
Troubleshooting 241