D70 Optics Error - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message

D70 Optics Error: 2

Qualifiers:
2 Zero Calibration
Drift or 2 Consecutive
Optical Clarity Drift
Failures
D70 Optics Error:
3, 4, 7, 11
Qualifiers:
3 tHb Analytical
Detector Signal Out
of Range
4 Excessive Noise for
tHb Analytical
Detector During Dark
Measurement
7 Dark Current Out of
Range for tHb
Analytical Detector
11 Neon Lamp
Failure
Rev. C
Probable Cause and Corrective Action
The system detects a failure in the CO-ox optical components.
The system can measure samples and report results for
NOTE:
parameters that are still available.
1. Perform a 2-point calibration.
2. If the parameter remains out of calibration, replace the
cartridges to make the parameter available for analysis.
Refer to Replacing the Measurement and Wash/Waste
Cartridges in Section 3 in the Rapidpoint 400 Series
Operator's Manual if you want to replace the cartridges.
3. If the message appears again, record the message and call
for technical assistance.
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
1. Repeat 2-point calibrations.
2. Replace the measurement cartridge.
The system detects a failure in the CO-ox optical components.
The system can measure samples and report results for
NOTE:
parameters that are still available.
1. Perform a 1-point calibration.
2. If the parameter fails the calibration, shut down the system
as described in Shutting Down the System, page 256, wait
10 seconds, and then turn the system on.
3. If the parameter remains out of calibration, call for technical
assistance.
(Continued)
Troubleshooting 205