Calibrating the System

Use this procedure to perform a 1-point, a 2-point, or a full calibration.
Only operators with level 1 or 2 security access can calibrate the system.
1. If prompted, enter your password, the service password, or the
password of the day.
2. At the Analysis screen, touch the
3. Touch
Calibrate
You cannot perform calibrations for at least 3 hours after installing
NOTE:
a new measurement cartridge. During this period, the system performs
automatic calibrations.
Some calibration types may not be available because an automatic
NOTE:
calibration is scheduled to begin shortly. For example, a 1-point calibration
is not available if a 2-point calibration is scheduled to start within
30 minutes.
Rev. C
.
button.
Status
Calibrations 143