This Password Is About To Expire. Renew The Password Before Access To The System Is Denied; This Password Is Expired. Renew The Password To Access The System; Uncorrected; Unrecoverable System Error - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message
This password is
about to expire.
Renew the password
before access to the
system is denied.
This password is
expired. Renew the
password to access
the system.

Uncorrected:

Unrecoverable
System Error.
Call your service
representative for
assistance.
228
Probable Cause and Corrective Action
This message may appear on systems connected to the
Rapidlink system. The message indicates that 14 days
remain before you exceed your certification date and will be
prevented from accessing the system. Contact your system
supervisor to renew your password.
This message may appear on systems connected to the
Rapidlink system. The message indicates that you have
exceeded your certification date and cannot access the
system. Contact your system supervisor to renew your
password.
The system did not correct the hematocrit (Hct) result for
+
either sodium (Na
) or potassium (K
potassium result is beyond the reporting range or because the
sodium or potassium sensor is out of calibration. The system
uses a default value of 140 mmol/L for sodium or a value of
4 mmol/L for potassium to correct the Hct result. On the
screen and the report, the letter u appears next to the result.
If the sodium or potassium result is beyond the reporting
range, analyze the sample again, if possible, or analyze a QC
sample, to verify the result.
If sodium or potassium is out of calibration, subsequent
calibrations may make the sensor available again. If sodium
or potassium remains out of calibration, replace the
cartridges as described in Replacing the Measurement and
Wash/Waste Cartridges in Section 3 in the Rapidpoint 400
Series Operator's Manual.
The system detects a problem that it cannot correct. Call for
technical assistance.
Service Solutions
View the events log for errors that may have caused the
problem and refer to the appropriate section of this manual.
+
) because the sodium or
(Continued)
Rev. C