Symbols on Screens and Reports

Refer to this section to identify the symbols that can appear with results on
screens and printed reports.
- - - - - ↑ or - - - - - ↓ These symbols appear instead of a result on the Results screen and the
Probable Cause
The result for the parameter is
above (- - - - -↑) or below (- - - - - ↓)
the reporting range for the
parameter.
- - - - -? This symbol appears instead of a result on the Results screen and the report..
Probable Cause
The system detected an atypical
response when measuring the
parameter.
When this symbol appears for the
hematocrit (Hct) result, it may
indicate that the Hct result was not
reported because sodium (Na
failed Required QC or AutomaticQC
analysis, or Required QC was not
performed.
162
report.
Corrective Action
1. Use the correct sample collection techniques and
2. Use the recommended storage, handling, and
3. Analyze the sample again ensuring that the sample
4. Analyze QC samples and verify that the QC results
Corrective Action
Analyze the sample again to verify the result.
If Required QC analysis was not performed for
sodium, perform the scheduled QC and then analyze
the sample again.
+
)
anticoagulant as described in Collecting Patient
Samples in Section 1 in the Rapidpoint 400 Series
Operator's Manual.
mixing techniques as described in Collecting
Patient Samples in Section 1 in the Rapidpoint 400
Series Operator's Manual.
has no bubbles. Checking for bubbles is especially
important for samples introduced from capillary
tubes.
are within the target ranges.