Sample Chamber Test - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

5. Touch the
The CO-ox lamp turns ON,
6. Touch the
The CO-ox lamp turns OFF.
The system automatically turns the lamp off when you exit the
NOTE:
Diagnostics screen.

Sample Chamber Test

Use the Sample Chamber test to verify that the sliding sample chamber
moves to the closed and open positions and to check the operation of the
sample chamber drive/interface and associated electronics in the sample
chamber interface assembly.
During run-time the sample chamber opens and closes to perform
NOTE:
the most optimum wash of the CO-ox optical chamber. An error flag
indicating a sample chamber position error is detected if the sample
chamber fails to move to the targeted position during testing.
A measurement cartridge or a measurement cartridge simulator
NOTE:
must be in place when this test is run.
1. Touch the
2. Touch
Sample Totals
The system cannot perform routine fluidic activities while you
CAUTION:
are at the Diagnostics screen or while you are performing diagnostic tests.
When you exit from the Diagnostics screen, the system may perform an
extended calibration before you can use the system.
3. Touch
Diagnostics
4. Enter the service password, or use the password of the day.
The Diagnostics screen appears.
5. Touch the
The second Diagnostic screen appears.
6. Touch the
Rev. C
CO-ox Lamp On/Off
CO-ox Lamp On/Off
button
Recall
.
.
.
Sample Chamber
, and
Start, Change
button.
button a second time.
button.
buttons sequentially to run the test
Stop
Troubleshooting 255