Unsuccessful Connection. Review The Setup; And That The Network Is Operating; W Cartridge Expired; Unsuccessful Connection - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message
Unsuccessful
Connection. Review
the setup values.
Ensure that the
cables are connected
and that the network
is operating.

W Cartridge Expired

Rev. C
Probable Cause and Corrective Action
When you touched the
screen, the system attempted to connect to the Rapidlink data
management system or the LIS, but the connection was not
successful.
1. Touch the
Continue
2. Verify that you entered the correct communication settings
and then touch the
Refer to Appendix E in the Rapidpoint 400 Series
Operator's Manual for more information about the
communication settings.
3. If the connection fails again, ensure that the Rapidlink
system or the LIS is able to accept messages from the
Rapidpoint 400 series system.
4. Ensure that the Rapidlink data management system or the
LIS is correctly configured to communicate with the
Rapidpoint 400 series system.
5. Ensure that the cable is not damaged and that it is the
correct cable for connecting the systems.
6. If necessary, touch
turn off the connection until you can resolve the problem.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed.
1. Ensure the communications setup information is correct.
2. Look at the LED at the network connector on the UIP board
to ensure the lights are on.
3. Check cables and connectors.
4. Replace the rear interconnect cable.
5. Replace the UIP board.
The wash/waste cartridge has exceeded its operating life or
is depleted.
Refer to Replacing the Wash/Waste Cartridge in Section 3 in
Rapidpoint 400 Series Operator's Manual to replace the
the
wash/waste cartridge.
button at the LIS Setup
Continue
button to display the LIS Setup screen.
button.
Continue
at the Communications screen to
None
Troubleshooting 229