D24 AQC Material
Error
D24 Qualifier Descriptions
A D24 error is detected while positioning, delivering or measuring AQC
material, and may be caused by the following conditions:
AutomaticQC material is depleted
Obstruction in the fluidic path of the Measurement or AutomaticQC
cartridge
Disconnected AutomaticQC cartridge connector or reagent bag tubing
Bubbles or short sample in the AutomaticQC material during
positioning to FD2
AutomaticQC material not detected at FD1
Obstruction in the fluidic path of the AutomaticQC cartridge
D24 errors are cleared by a successful AutomaticQC cartridge
replacement or by turning off the AQC feature in Setup.
192
Probable Cause and Corrective Action
The system detects that the flow of one or more quality control
materials is inadequate or incorrect. The system
automatically repeats the QC sample. If the system cannot
correct the problem, it prompts you to replace the
AutomaticQC cartridge.
1. If prompted, replace the cartridge as indicated in the
message. Replacing the AutomaticQC Cartridge on
page 168.
2. If the message appears again, record the message and call
for technical assistance.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed.
1. Remove the AutomaticQC cartridge and perform the AQC
valve test and observe the valve moving.
2. Use the AutomaticQC cartridge simulator to test the valve
alignment. Refer to Using the AutomaticQC Cartridge
Simulator on page 246 for information on the cartridge
simulator.
3. Replace the AQC frame assembly.