Restore Setup and Correct the Date and Time

1. Touch
Cancel
cartridge change.
2. Touch
Setup
3. Enter the default password,
4. Touch
Secured Options.
5. Touch the down arrow and then touch
6. When prompted, insert the Setup diskette into the diskette drive and
then touch the
The wait screen appears while the data is copied to the system.
7. When prompted, remove the diskette from the drive and touch the
Continue
8. Touch
System Options
9. Touch
Date and Time
10. Change the date and time and then touch the
11. Leave the Rapidpoint 400 series system on the Setup screen and
continue on to Rapidlink Setup.
After the Rapidpoint Setup is restored, the Rapidlink system must
NOTE:
be set up to communicate with the Rapidpoint 400 series system.

Rapidlink Setup

1. At the Rapidlink workstation, access the Device Options screen:
a. Select
b. Select
c. Select
d. Select the existing Rapidpoint 400 series system that is being
swapped out and change the Communication Mode from
CompleNet
Write down which Auto Save options are chosen for the
NOTE:
Rapidpoint 400 series system.
Rev. C
when prompted for the measurement and wash/waste
.
button.
Continue
button.
.
.
.
Setup
.
Continue
.
Devices
to
.
Offline
Swapping Out a Rapidpoint 400 Series System 415
, and then touch the
12345
Restore Setup
button.
Continue
.
button.
Continue