d. View the video before you begin if required.
e. Push in and then slide the connector on the AutomaticQC cartridge
to the right.
f.
Wait for the AutomaticQC cartridge to eject from the system.
g. Remove the AutomaticQC cartridge and dispose of this cartridge.
Wait 10 minutes until the Return button appears on the Replacing
the AutomaticQC Cartridge screen.
h. Touch the
i.
A message appears indicating that the cartridge needs replacing.
j.
Touch
The Status screen appears.
4. Turn off the AutomaticQC option:
a. Touch
b. Touch
c. Touch
d. Touch the
The Status screen appears.
Clean and disinfect any areas that may be contaminated from spills
NOTE:
around the measurement and wash/waste cartridges. Refer to Cleaning
and Disinfecting the Exterior Surfaces in the Rapidpoint 400 Series
Operator's Manual.
5. Remove the measurement and wash/waste cartridges and dispose of
them according to the institution's protocol for disposal of
biohazardous materials. Refer to Replacing the Measurement and
Wash/Waste Cartridges in Section 3 in the Rapidpoint 400 Series
Operator's Manual, if required.
a. Touch the
A message appears message indicating that the cartridge does
not need replacing.
398
button.
Return
.
Cancel
.
Setup
.
QC Options
.
Unscheduled QC
button twice.
Continue
Measurement Cartridge
button and then touch
.
Replace
Rev. C