Output Signal
REF_SIG
HCT1_SIG
HCT2_SIG
THM_BLK_HI
THM_BLK_LO
THM_PRE_HI
THM_PRE_LO
THM_SMP_HI
THM_SMP_LO
PO2_RF
PO2_CN
GLU_RF
GLU_CN
LAC_RF
LAC_CN
CRE_RF
CRE_CN
HCT1_HS
HCT1_LS
HCT2_HS
HCT2_LS
Rev. C
Description
Amplified and buffered reference sensor signal.
Amplified and buffered hematocrit 1 sensor signal.
Amplified and buffered hematocrit 2 sensor signal.
The pass-through signal connected to one end of the block
thermistor.
The pass-through signal connected to the other end of the
block thermistor.
The pass-through signal connected to one end of the
preheater thermistor.
The pass-through signal connected the other end of the
preheater thermistor.
The pass-through signal connected to one end of the sample
thermistor.
The pass-through signal connected to the other end of the
sample thermistor.
The voltage sensed internal to the pO
The voltage control electrode for the pO
The voltage sensed internal to the glucose sensor. The
voltage is maintained at -400 mV ±4% and constant ±1 mV
between calibration cycles with respect to 5_RTN_GLU.
The voltage control electrode for the glucose sensor. The
voltage adjusts such that the signal at GLU_RF is -400 mV
with respect to 5_RTN_GLU.
For future use.
For future use.
For future use.
For future use.
The excitation signal for the hematocrit 1 sensor.
Signal return for the hematocrit 1 sensor.
The excitation signal for the hematocrit 2 sensor.
Signal return for the hematocrit 2 sensor.
sensor.
2
sensor.
2
Electronic System Overview 73
(Continued)