AutomaticQC material not detected at FD1
•
Obstruction in the fluidic path of the measurement or AutomaticQC
cartridge
A D41 is detected during the priming of an AutomaticQC cartridge. D41 is
cleared by a successful AutomaticQC cartridge replacement, turning off
AQC in Setup, or replacing both the measurement and waste/wash
cartridges.
Rev. C
Troubleshooting 203