System Error. Turn The Power Switch Off; Wait 10 Seconds Then Turn The Switch On; System Requires Operator Attention; Switch Off - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message
System Error.
Turn the power
switch off. Wait
10 seconds then
turn the switch on.
System requires
operator attention.
Rev. C
Probable Cause and Corrective Action
An electronic or processing error has occurred.
1. Shut down the system as described in Shutting Down the
System on page 256, wait 10 seconds, and then turn the
system on.
2. If the message appears again, record the message and call
for technical assistance.
Service Solutions
View the events log for errors that may be causing the problem
and refer to the appropriate section of this manual.
The message appears on the Sign-In screen if any of the
following conditions occurs:
Required QC analysis is currently scheduled
The cartridges are nearly expired or depleted.
One or more parameters are turned off because they failed
Required QC or AutomaticQC analysis, Required QC
analysis was not performed when scheduled, or parameters
failed calibration.
1. Enter your password and touch the
2. View the Analysis screen to determine what action is
required and then take the appropriate action.
Refer to Button and Parameter Status, Section 5 in the
Rapidpoint 400 Series Operator's Manual, if you need more
information about the status of buttons at the Analysis
screen.
Refer to Section 5, Understanding the Rapidpoint 400
Series System in the Rapidpoint 400 Series Operator's
Manual, if you need more information about the Analysis
screen and the symbols that appear at this screen and on the
banner.
button.
Continue
(Continued)
Troubleshooting 221