3. Disconnect the patch cord from the in-line coupler on the rear
interconnect board.
1 Patch cord
2 In-line coupler
3 Rear interconnect board
4 Rear interconnect cable
4. Remove the rear interconnect board:
The rear interconnect cable is soldered to the board.
NOTE:
a. Remove the four hex standoffs from the frame.
b. Remove the screw holding the clip at the top of the board.
5. Remove the in-line coupler from the board.
6. Install the new rear interconnect board, and reinstall the in-line coupler.