Testing the IDEE ROM and Cartridge Connector Switch.

Ensure that the AutomaticQC option is turned on in Setup.
NOTE:
If you want to . . .
test the IDEE ROM and the cartridge
connector switch
complete the valve alignment test
15. Touch the Continue button until you return to the Status screen.
The message informing you that the AQC cartridge needs replacing
appears.
16. Touch Cancel to remove the message.
17. View the messages in the Event Log on the Status screen.
If . . .
the AQC Cartridge Not Valid message
appears
the No AQC Cartridge message appears the system cannot read the IDEE
18. If the system is able to read the IDEE ROM, test the cartridge connector
switch.
a. Move the switch on the cartridge simulator to the OFF position and
view the Events Log.
If . . .
the AQC Connector is Open
message appears
this message does not appear
b. When the switch is moved back to the ON position, the message
disappears.
19. Return to the Diagnostics screen.
20. When the test is complete, put the AQC cartridge simulator switch in
the OFF position.
Be careful to prevent the simulator from being ejected clear of
CAUTION:
the system.
Rev. C
Then . . .
continue with step 15.
go to step 19.
Then . . .
the system is able to read the IDEE
ROM.
ROM.
Then . . .
the system is able to see the
connector switch.
the system is unable to see the
connector switch.
Troubleshooting 251