Replacing the Door Lock Module

BIOHAZARD:
Observe precautions for handling electrostatic-sensitive
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Tools and Supplies
Phillips screwdriver, #1
small flat-blade screwdriver
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the door lock module:
a. Slide the front of the system to the edge of the work surface.
Using a small flat-blade screwdriver, access the door lock solenoid
through the middle hole of the three access holes in the bottom
front right corner of the system and push the solenoid toward the
back of the system until the front door opens.
Remove the two screws securing the door lock module.
b. Slide the door lock module from the front of the interface wall.
c. Disconnect the door switch connector.
3. Install the new door lock module.
Rev. C
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.