Table of Contents

Advertisement

2. Check the measurement cartridge valve alignment:
a. Remove the measurement and the wash/waste cartridges. Refer
to Replacing the Measurement and Wash/Waste Cartridges in
Section 3 in the Rapidpoint 400 Series Operator's Manual.
NOTE:
tighten the screw to hold the valve alignment gauge in place.
b. Install the valve alignment gauge, and hold it in position.
3. Touch the
4. Touch
Sample Totals
5. Touch
Diagnostics
6. Enter the service password, or use the password of the day.
7. Touch the
8. Touch
M Valve
9. Touch
Start.
10. Touch
Change.
11. With the valve in position 2, verify that the gauge reads within 0 ± 5.
12. Touch
Stop
13. Wait until the cartridge interface returns, then remove the valve
alignment gauge.

Using the AutomaticQC Cartridge Simulator

BIOHAZARD:
recommended precautions when working with biohazardous materials.
Use the AutomaticQC cartridge simulator along with the diagnostics
software to diagnose problems in valve alignment, and with the IDEE ROM
and cartridge connector switch.
The cartridge simulator is powered by the Rapidpoint 400 series system
and requires no maintenance.
246
If the system is configured to receive the thumb screw,
button.
Recall
.
.
button.
Continue
.
The valve moves to position 1 on the screen.
The valve moves to position 2 on the screen.
when done.
Refer to Protecting Yourself from Biohazards for
Rev. C