Not in Retro Calibration Mode
If the measurement module temperature is outside the range 37 ± 0.20°C
the temp warning exception is set. This exception does not indicate failure
of the temperature control system. Rather, it alerts the user that the
instrument is in the process of warming up and is close to the normal
operating temperature. Sample analysis requests will be accepted. This
exception is generated if the temperature control software was down for a
few minutes or more.
As soon as the temperature gets within the range 37 ± 0.20°C, the Temp
Warning message disappears.
All Modes
Starting sample operations while this exception is posted may generate the
Temp Out of Range exception.
Refer to the related exceptions: Temp Not Ready and Temp Out of Range.
If during a non-fatal temperature error no measurement cartridge is
installed then the measurement cartridge replacement procedure will
preempt the temperature handling.
Message
The language cannot
be selected because
the current version is
not installed. Install
the latest language
version to select the
language.
226
Probable Cause and Corrective Action
The current version for the language selected in Setup is not
installed on the system.
1. Obtain the software that contains the latest version of the
language that you want to select. It should be the same
version as the English language installed on the system.
2. Install the latest version for the language that you need.
Refer to Installing New System Software in Section 7 i n the
Rapidpoint 400 Series Operator's Manual
3. Select the language again in Setup.
(Continued)
Rev. C