Bayer HealthCare RapidPoint 400 Series Service Manual page 243

Table of Contents

Advertisement

1. Remove the measurement and the wash/waste cartridges. Refer to
Replacing the Measurement and Wash/Waste Cartridges in Section 3
in the Rapidpoint 400 Series Operator's Manual.
2. Connect the Tygon tubing to the cartridge simulator.
1 Wash In
2 Wash Out
3 Sample Out
4 Sample In
3. Install the cartridge simulator.
Perform steps 4 through 14 to test the pump flow for either the
NOTE:
sample or the wash pump.
4. Place the inlet tubes into a beaker of reagent water.
5. Place the outlet tubes into an empty beaker.
If you are already in Diagnostics, go to step 11.
6. Touch the
7. Touch
Sample Totals
8. Touch
Diagnostics
9. Enter the service password, or use the password of the day.
10. Touch the
Rev. C
button.
Recall
.
.
button.
Continue
Troubleshooting 243