Replacing The Illumination Housing Assembly - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Replacing the Illumination Housing Assembly

BIOHAZARD:
Tools and Supplies
Phillips screwdriver, #1
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
3. Remove the illumination housing assembly:
CAUTION:
the optical fibers resulting in erroneous optics readings.
CAUTION:
the defusing material at each end of the cable as this can result in
erroneous optics readings.
a. Disconnect the fiber bundle from the illumination housing
assembly.
b. Using a Phillips screwdriver, remove the two Phillips screws,
located at the rear of the frame securing the illumination housing
assembly and retain.
360
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Do not bend, twist, or crimp the fiber bundle as this can damage
Handle the fiber bundle with care being careful not to damage
Rev. C