D33 Qualifier Descriptions
D33 Valve Error: 2 is triggered if the AQC cartridge does not eject in the
absence of a valve drive failure, and if the valve fails to move to a certain
position due to a drive valve failure or a latch failure.
When the AQC cartridge does not eject, the D33 is detected:
after closing the cartridge connector when replacing the cartridge
when the cartridge fails to eject from the system
during power up
When the valve fails to move, the D33 is detected:
during AutomaticQC
during the AQC Valve diagnostic test
during a smart AQC prime
during power up
Rev. C
Troubleshooting 195