Replacing The Hard Drive - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Replacing the Hard Drive

BIOHAZARD:
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Tools and Supplies
Allen wrench, 5/64
Phillips screwdriver, #1
CAUTION:
recalibrated. Refer to Calibrations, for instructions on calibrating the touch
screen.
Before you begin, record the information from the Service Data
NOTE:
screen and the IP addresses from the Communications screen if required,
and save the Setup data. Refer to Section 6 in the Rapidpoint 400
Operator's Manual to save the Setup data.
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
Retain grommets and nylon spacers after removing the hard drive.
NOTE:
3. Remove the hard drive:
a. Disconnect the hard drive cable from the hard drive and retain the
cable.
334
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Observe precautions for handling electrostatic-sensitive
After replacing the hard drive, the touch screen must be
Series
Rev. C

Advertisement

Table of Contents
loading

Table of Contents