D38 Temp Error: 1 - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message

D38 Temp Error: 1

Qualifiers:
1 Temperature
over range
D38 Temp Error: 2–13
Qualifiers:
2 Sample Temperature
Sensor Failure:
Thermistor is shorted or
open. Temperature is
<10°C or > 50°C.
198
Probable Cause and Corrective Action
An error has occurred in the temperature control system
because of a problem with the fan.
1. Ensure that the fan is working and nothing is blocking the
flow of air.
2. Inspect the air filter and replace it if required. Refer to
Replacing the Air Filter in Section 3 in the
Rapidpoint 400 Series Operator's Manual.
3. Shut down the system and then restart is as described
in Shutting Down the System on page 256.
4. If prompted to call for service, record the message
including the qualifier, and call for technical assistance.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed.
1. Check that the filter is not blocked or that nothing is
blocking air to the main board.
2. Check cables and connections to the fan.
3. Check that the fan is operating, and replace, if required.
An error has occurred in the temperature control system.
This error can occur if a component in the temperature
control system has failed. The error can also occur if you
operate the system at ambient temperatures beyond the
operating temperature.
1. Wait while the system attempts to correct the problem.
2. Operate the system in an environment with ambient
temperature within 15 to 32
3. If prompted, shut down the system and then restart it as
described in Shutting Down the System on page 256.
°
C.
(Continued)
Rev. C