Ca++, Cl-, Hct; D4 Offset Error: Glu, Po2 - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message
D4 Offset Error:
pH, pCO
, Na
2
++
-
Ca
, Cl
, Hct
D4 Offset Error:
Glu, pO
2,
186
Probable Cause and Corrective Action
The system turned the parameter identified in the message
+
+
, K
,
off because the sensor exceeded calibration limits.
1. Perform a 1-point calibration.
2. If the parameter fails the calibration, wait as subsequent
calibrations may make the parameter available again.
3. If the parameter remains out of calibration, replace the
cartridges to make the parameter available for analysis.
Refer to Replacing the Measurement and Wash/Waste
Cartridges in Section 3 in the Rapidpoint 400 Series
Operator's Manual if you want to replace the cartridges.
4. If the message appears again, record the message and call
for technical assistance.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed. If cartridge replacement does
not solve the problem:
1. Use the measurement cartridge simulator to perform the
sensor test.
2. Replace the connector block.
3. Replace the cartridge interface module.
The system turned the parameter identified in the message
off because the sensor exceeded calibration limits.
1. Perform a 2-point calibration.
2. If the parameter fails the calibration, wait as subsequent
calibrations may make the parameter available again.
3. If the parameter remains out of calibration, replace the
cartridges to make the parameter available for analysis.
Refer to Replacing the Measurement and Wash/Waste
Cartridges in Section 3 in the Rapidpoint 400 Series
Operator's Manual if you want to replace the cartridges.
4. If the message appears again, record the message and call
for technical assistance.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed. If cartridge replacement does
not solve the problem:
1. Use the measurement cartridge simulator to perform the
sensor test.
2. Replace the connector block.
3. Replace the cartridge interface module.