D33 Valve Error: 2 - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message

D33 Valve Error: 2

Qualifiers:
2 AutomaticQC
cartridge
194
Probable Cause and Corrective Action
The system detects a problem with the valve inside the
AutomaticQC cartridge either during cartridge replacement
or during sample analysis.
If the problem occurs while trying to replace the AutomaticQC
cartridge:
1. Check that the connector is completely open and then touch
Replace to try and eject the cartridge.
2. If the message appears again, record the message,
including the qualifier, and call for technical assistance.
If the problem occurs during AutomaticQC analysis:
1. Shut down the system as described in Shutting Down the
System on page 256, wait 10 seconds, and then turn the
system on.
2. If the message appears again, record the message,
including the qualifier, and call for technical assistance.
Service Solutions
Perform service solutions only after all other corrective
actions have been completed.
1. Remove the AutomaticQC cartridge and perform the AQC
valve test and observe the valve moving.
2. Use the AutomaticQC cartridge simulator to test the valve
alignment. Refer to Using the AutomaticQC Cartridge
Simulator on page 246 for information on the cartridge
simulator.
3. Replace the AQC frame assembly.
Rev. C