The Replace Cartridges symbol appears in the banner.
Probable Cause
The measurement and wash/waste
cartridges will soon be depleted or
expired. The banner shows the
number of samples or amount of
time (00:34 minutes in this example)
remaining until the cartridges are
depleted or expired.
The Wash/Waste Cartridge symbol appears in the banner.
Probable Cause
The wash/waste cartridge will soon
be depleted or expired. The banner
shows the number of samples or
amount of time (00:34 minutes in this
example) remaining until the wash/
waste cartridge is depleted or
expired.
The AutomaticQC Cartridge symbol appears in the banner.
Probable Cause
The AutomaticQC cartridge will soon
be depleted or expired. The banner
shows the number of samples or
amount of time (00:34 minutes in
this example) remaining until the
Automatic QC cartridge is depleted
or expired.
160
Corrective Action
No action is necessary at this time. You can monitor
the number of samples and amount of time remaining
to determine when you want to replace the
cartridges. The system prompts you to replace the
cartridges when they are depleted or expired.
Corrective Action
No action is necessary at this time. You can monitor
the number of samples and amount of time remaining
to determine when you want to replace the wash/
waste cartridge. The system prompts you to replace
the cartridge when it is depleted or expired.
Corrective Action
No action is necessary at this time. You can monitor
the number of samples and amount of time remaining
to determine when you want to replace the Automatic
QC cartridge. The system prompts you to replace the
cartridge when it is depleted or expired.