The preheater chip is located in front of the measurement block. The
sample passes through the preheater during aspiration. The preheater
temperature range is 37.0 to 37.6°C, with a target preheater temperature
of 37.3°C.
The IDEE ROM chip informs the system that the correct measurement
cartridge is installed and helps in diagnosing cartridge problems. The
IDEE Rom Identifies the following:
analyzer software version
date installed on the analyzer
date removed fro analyzer (if removed normally)
analyzer serial number

KCl Reservoir

The KCl reservoir continuously refreshes the KCl concentration on the
reference sensor. Each measurement cartridge includes a KCl vial
containing 16 mL of liquid KCl.
The KCl reservoir is clamped closed to prevent it from leaking during
shipment. The tubing on the KCl vial is automatically unclamped allowing
the KCl to flow when the measurement cartridge is installed.